Radlink CR Pro Installationsanleitung

Stöbern Sie online oder laden Sie Installationsanleitung nach Stromverteilungseinheiten (PDUs) Radlink CR Pro herunter. Radlink CR Pro Installation guide Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Setup and Installation Guide

00.09-002A CR-Pro Computed Radiology Imaging Device Setup and Installation Guide April 2010 Radlink, Inc. 2400 Marine Ave. Redondo Be

Seite 2

9 Hardware Installation 1. Place the CR-Pro on a solid flat, leveled surface. 2. Plug the power cord to the AC Power Adapter located on the bac

Seite 3 - Preface

10 Chapter 2 Setup Guide and Configuration

Seite 4 - LASER SAFETY –

11 STEP 1: Application Start Up 1. To start the application double click the Radlink Icon located on your desktop. (Figure 1) Fig

Seite 5

12 2. Image Pro software home screen will appear as Figure 2.Figure 2

Seite 6

13 3. Verify that Scanner Status is lit green with READY. (Figure 3) Figure 3

Seite 7

14 STEP 2: Calibration Procedure Calibration of the CR PRO is required to compensate for any hardware differences and should be performed at least on

Seite 8 - Introduction

15 4. Start Calibration From the application Click Manage  CR Setup  Calibrate. (Figure 5) The calibration process should take 3-4 minutes to c

Seite 9

16 Chapter 3 Test Scan Procedure

Seite 10 - Hardware Installation

17 STEP 1: Expose an Image You are now ready to scan your first cassette. Expose an image using one of the cassette(s) that you purchased.

Seite 11 - Setup Guide

18 STEP 3: Scan Cassette Select New Patient to add or New Study to add an existing patient. For detailed description of the functions and operatio

Seite 12 - Figure 1

1 © Copyright 2008 by Radlink, Inc All rights reserved Printed in USA Specifications and product and/or service offerings are subject to change wi

Seite 13 - Figure 2

19 STEP 4: Patient Info Window Patient Information window will display, input the following: ID, Sex, Last Name, First Name, Middle Name and Birth

Seite 15 - Figure 4

21 STEP 5: Selecting Body Part 1. Input the same technique used when you expose the cassette. 2. Click Save 3. Click Scan CR bu

Seite 16 - Figure 5

22 STEP 6: Scan Cassette The CR Pro will now begin digitizing your exposed image.

Seite 17 - Chapter 3

23 STEP 8: Complete Study A sample scan of a knee phantom is shown below. Fig

Seite 18 - STEP 2: Startup Window

24 Chapter 4 Troubleshooting Guide

Seite 19 - STEP 3: Scan Cassette

25 General Troubleshooting Problem Action No power to the CR Pro Ensure the monitor power is on. Check that the Power Switch on the back of the C

Seite 20 - STEP 4: Patient Info Window

26 Cassette Related Problems Troubleshooting Symptom Possible Causes What To Do Damage to outer enclosure • Rough handling by technicians • Patien

Seite 21

27 Chapter 5 Maintenance Procedures

Seite 22 - STEP 5: Selecting Body Part

28 CR ProTM Phosphor Plate Cassette Maintenance Procedure The Phosphor Plate Imaging Cassette is a key component in the CR ProTM Comput

Seite 23 - STEP 7: Auto Erasing

2 Preface • THANK YOU – For purchasing the Radlink CR-Pro Computed Radiology Imaging Device. This manual will assist you in learning all the fun

Seite 24 - STEP 8: Complete Study

29 Care and Cleaning To ensure optimum performance from the cassettes and to eliminate potential image artifacts, it is recommended that the casset

Seite 25 - Troubleshooting Guide

30 4. Twist the screwdriver to extract the phosphor plate (see photo #2). To prevent damaging the phosphor plate, do not apply pressure to the cente

Seite 26 - Problem Action

31 7. Wipe the long edges of the plate to remove any accumulated residue (see photo #3). Cleaning the edges of the plate will reduce the possibilit

Seite 27

32 10. Wipe the surface of the phosphor plate in the long direction with overlapping stokes. Rotate the cleaning cloth to use a clean section as you

Seite 28 - Chapter 5

33 15. Turn the cassette over and clean the lower corners where the side and bottom edges of the cassette join together (see photo #5). Photo 5 1

Seite 29

34 Appendix

Seite 30

35 CR-Pro Specifications . Table 1 –System Hardware Specifications 2100 Series Specification 2200 Series Specification Processor Intel Pentium

Seite 31

36 Table 2 – Hardware Specifications Phosphor Plate Phosphor Plate 8” X 10” (203.2mm X 254.0mm) 10” X 12” (254.0mm X 304.8mm) 14” X 17” (355.6mm

Seite 32

37 USER QUALITY ASSURANCE & MAINTENANCE PROGRAM FOR THE CR-PRO INTRODUCTION: Understanding the basic fundamentals and automating the decisions

Seite 33

38 Chart to define the variables for each X-Ray. Or, better yet, if the site’s Techniques Chart is known by Radlink prior to shipment, a customized

Seite 34

3 • SAFETY – Read and follow all warning and safety instructions in this User’s Manual and marked on the product. Not following the instructions may

Seite 35 - Appendix

39 Radlink Inc. CR-Pro Computed Radiology Imaging Device DICOM Conformance Statement 1. Introduction This document provides conformance by the Ra

Seite 36 - CR-Pro Specifications

40 2. Implementation Model The CR Pro is a medical acquisition system that supports Computed Radiography (CR). The CR Pro encompasses the following

Seite 37

41 Figure 1: Image Transfer Data Flow Diagram 2.1.2 Functional Definitions of Application Entities The CR Pro Image Tra

Seite 38 - PROGRAM FOR THE CR-PRO

42 2.2 Modality Worklist 2.2.1 Application Data Flow Figure 2: Modality Worklist Data Flow Diagram 2.2.2 Functional De

Seite 39

43 3. Radlink CR Pro Image Transfer AE Specifications 3.1 CR Pro Specifications 3.1.1 SCU Verification This AE provides standard conformance

Seite 40 - DICOM Conformance Statement

44 3.2.4 Implementation Identifying Information The CR Pro Image Transfer AE is identified by the following ids: Implementation Class UID: 1.2.250.

Seite 41 - Implementation Model

45 3.3.3 SOP Specific Conformance for Non-Compressed Image Storage Table 5 illustrates the encoding for a Non-Compressed Image. Table 5: Non-Comp

Seite 42 - Standard Interface

46 Series Number (0020,0010) IS nnnnn Image Number (0020,0013) IS nnnnn Samples per Pixel (0028,0002) US 1 Photometric Interpretation (0028,000

Seite 43

47 4.1.2 Number of Associations The Radlink CR Pro MWL AE initiates one association with the default remote Modality Worklist SCP. The association

Seite 44

48 4.2.2 SOP Specific Conformance If the Radlink View Pro Modality Worklist component is unable to open an association with the selected destinatio

Seite 45

4 Electromagnetic Interference (EMI) is any signal or emission, radiated in free space or conducted along power or signal leads, that endangers the f

Seite 46

49 4.3 Association Acceptance Policy The Radlink View Pro DICOM Modality Worklist Component does not accept associations. 5 Communication Pr

Seite 47

50 9 Abbreviations used in the Value Representation (VR) Column of Supported DICOM Data Elements Code Name Meaning* Length AE Application entity

Seite 48

51 TM Time A time string hhmmss.ffffff with shortcut possibilities 16 Bytes Maximum UI Unique identifier A string with one identifier consisting

Seite 49

5 The AC outlet should be dedicated to the CR-Pro (no other devices connected to this circuit) to prevent the possibility of injected line “noise “by

Seite 50

6 TABLE OF CONTENTS Chapter 1: Hardware Installation Chapter 2: Setup and Configuration Chapter 3: Test Scan Guide Chapter 4: Troubleshoo

Seite 51 - Meaning* Length

7 Introduction The Radlink CR-Pro is a radiological device, which digitizes erasable phosphor plates and transmits the image to an onboard compute

Seite 52

8 Chapter 1 Hardware Installation

Kommentare zu diesen Handbüchern

Keine Kommentare